TOWEL OR 17X24IN BLUE STERILE 6/PK 28700-006

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2019-03-21 for TOWEL OR 17X24IN BLUE STERILE 6/PK 28700-006 manufactured by Jianerkang Medical Dressing Co.,.

Event Text Entries

[139439098] Based on the supplier investigation, the device history record review did not indicate any exception that could lead to the reported incident. Or towels are made of cotton, so lint is born and inevitable. The average linting data is 0. 120/ 10 pieces. Six pieces of samples were returned for investigation. There was no lint falling from the sample surfaces. The supplier conducted linting data tests and the linting data is 0. 106g / 6 pieces, which meets the acceptable criteria (=0. 38g/10 pieces). From the investigation, nothing abnormal happened during production and all linting test data was within the acceptable range. Therefore, the root cause for the reported incident could not be determined. The complaint information was provided to the relevant sectors for their awareness. No action will be taken at this time, but we will continue to monitor the trend of this type of incident.
Patient Sequence No: 1, Text Type: N, H10


[139439099] We had an issue with some sterile blue towels. We were performing an aneurysm coiling case. The wires are laid upon the wet or towels and the wires are then placed into the flush bowl. The sterile blue towels had fuzz come off into our flush bowl. That flush bowl is used to flush the catheters that go into the patient? S brain. There was no injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423537-2019-00292
MDR Report Key8439766
Report SourceCOMPANY REPRESENTATIVE
Date Received2019-03-21
Date of Report2019-03-21
Date of Event2019-02-21
Date Facility Aware2019-02-21
Date Added to Maude2019-03-21
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPATRICIA TUCKER
Manufacturer Street3651 BIRCHWOOD DR.
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478874151
Manufacturer G1JIANERKANG MEDICAL DRESSING CO
Manufacturer StreetNO.1, JIANERKANG ROAD ZHIXI TOWN JINTAN CITY
Manufacturer CityJIANGSU 213251
Manufacturer CountryCH
Manufacturer Postal Code213251
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOWEL OR 17X24IN BLUE STERILE 6/PK
Generic NameTOWEL OR 17X24IN BLUE STERILE 6/PK
Product CodePUI
Date Received2019-03-21
Returned To Mfg2019-03-04
Model Number28700-006
Catalog Number28700-006
Lot Number181006-18-SH
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerJIANERKANG MEDICAL DRESSING CO.,
Manufacturer AddressNO.1, JIANERKANG ROAD ZHIXI TOWN JINTAN CITY JIANGSU 213251 CH 213251


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-21

© 2026 FDA.report