WALLFLEX BILIARY M00570370 7037

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-03-21 for WALLFLEX BILIARY M00570370 7037 manufactured by Boston Scientific Corporation.

Event Text Entries

[139530746] The complainant was unable to report the suspect device lot number; therefore, the manufacture and expiration dates are unknown. However, it was reported the device was not used past its expiry date. (b)(4). The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed. If any further relevant information is identified, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[139530747] It was reported to boston scientific corporation on (b)(6) 2019 that a 10 x 60 wallflex biliary rx fully covered rmv stent was to be used to treat a malignant stricture in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2019. Reportedly, the patient's anatomy was slightly tortuous and was dilated prior to procedure. According to the complainant, during the procedure, the stent was able to be deployed; however, during catheter removal, the delivery system got caught on the stent and was difficult to remove. Reportedly, the physician tried to release the delivery system from the stent but ended up pulling the stent out of position into the duodenum. The stent was removed with grasping forceps and a 10 x 80 wallflex biliary stent was used to complete the procedure. There were no patient complications reported as a result of this event. The patient's condition following the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2019-01312
MDR Report Key8441496
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-03-21
Date of Report2019-03-21
Date of Event2019-02-26
Date Mfgr Received2019-02-26
Date Added to Maude2019-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWALLFLEX BILIARY
Generic NameBILIARY STENT SYSTEM FOR BENIGN STRICTURES
Product CodePNB
Date Received2019-03-21
Model NumberM00570370
Catalog Number7037
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-03-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.