MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2019-03-21 for 3/16" (4.8MM) X 6' (1.8M) SURGICAL TUBING 0036480 manufactured by Consolidated Medical Equipment.
[139543833]
At time of filing, although expected, the reported device has not been returned to conmed for evaluation. A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation. This issue will continue to be monitored through the complaint system to assure patient safety.
Patient Sequence No: 1, Text Type: N, H10
[139543834]
During incoming inspection, the distributor rejected this device, 0036480, lot 201802234, for a possible insufficient heatseal. There was no contact with the patient as this was found during incoming inspection. Due to the potential severity of a breach in sterility, this complaint meets the criteria for a reportable event. This report is being raised as a malfunction with potential for injury upon reoccurrence.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3007305485-2019-00078 |
MDR Report Key | 8441637 |
Report Source | DISTRIBUTOR,FOREIGN |
Date Received | 2019-03-21 |
Date of Report | 2019-04-18 |
Date Mfgr Received | 2019-04-05 |
Device Manufacturer Date | 2018-02-23 |
Date Added to Maude | 2019-03-21 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS TRACEY WEISELBENTON |
Manufacturer Street | 11311 CONCEPT BLVD |
Manufacturer City | LARGO FL 337734908 |
Manufacturer Country | US |
Manufacturer Postal | 337734908 |
Manufacturer Phone | 7273995557 |
Manufacturer G1 | CONSOLIDATED MEDICAL EQUIPMENT |
Manufacturer Street | ALEJANDRO DUMAS AVE, 11321 COMPLEJO INDUSTRIAL CHIHUAHUA |
Manufacturer City | CHIHUAHUA, MEXICO 31136 |
Manufacturer Country | MX |
Manufacturer Postal Code | 31136 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 3/16" (4.8MM) X 6' (1.8M) SURGICAL TUBING |
Generic Name | TUBE, ASPIRATING, FLEXIBLE, CONNECTING |
Product Code | BYY |
Date Received | 2019-03-21 |
Catalog Number | 0036480 |
Lot Number | 201802234 |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CONSOLIDATED MEDICAL EQUIPMENT |
Manufacturer Address | ALEJANDRO DUMAS AVE, 11321 COMPLEJO INDUSTRIAL CHIHUAHUA CHIHUAHUA, MEXICO 31136 MX 31136 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-03-21 |