[139541846] (b)(4). The complainant indicated that the delivery system was disposed and will not be returned for evaluation, and the stent remains implanted. Therefore, a failure analysis of the complaint device could not be completed. If any further relevant information is identified, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[139541847] It was reported to boston scientific corporation on (b)(6) 2019 that a wallflex biliary rx fully covered rmv stent has been implanted in the common bile duct (cbd) to treat malignancy during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2019. Reportedly, the patient's anatomy was not tortous and was not dilated prior to stent placement. According to the complainant, during the procedure, the stent was able to be deployed; however, during catheter removal, the tip of the delivery system got caught on the proximal portion of the stent. Eventually, the delivery system was able to be pulled out and the stent was left implanted to complete this procedure. There were no patient complications as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2019-01320
MDR Report Key8441976
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-03-21
Date of Report2019-03-21
Date of Event2019-02-26
Date Mfgr Received2019-02-27
Device Manufacturer Date2018-09-10
Date Added to Maude2019-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWALLFLEX BILIARY
Generic NameBILIARY STENT SYSTEM FOR BENIGN STRICTURES
Product CodePNB
Date Received2019-03-21
Model NumberM00570370
Catalog Number7037
Lot Number0022621163
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.