[139711752] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[139711753] The initial reporter questioned results for 37 patient samples tested for cobas elecsys anti-tpo (anti-tpo) on a cobas 6000 e 601 module. The customer repeated 37 patient samples that had originally been released from the laboratory. Of that data, the results for 6 patient samples were discrepant. The customer is not sure which result is correct. There was no allegation that an adverse event occurred. The e601 module serial number is (b)(4). Qc results show a large scatter. Calibration data also shows large variations in signals. Sample pipetting error messages were observed on the alarm trace data from (b)(6) 2019. The measuring cell had been replaced during the annual preventive maintenance visit on 23-jan-2019. The customer stated they had water quality issues recently. The water filters were replaced on 18-mar-2019.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01181
MDR Report Key8443566
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-03-22
Date of Report2019-04-09
Date of Event2019-03-04
Date Mfgr Received2019-03-07
Date Added to Maude2019-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
Generic NameSYSTEM, TEST, THYROID AUTOANTIBODY
Product CodeJZO
Date Received2019-03-22
Model NumberNA
Catalog Number06368590190
Lot Number34258403
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-22