MPS 2 CONSOLE 5201260

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-22 for MPS 2 CONSOLE 5201260 manufactured by Quest Medical, Inc..

Event Text Entries

[139822986] The device was evaluated and he reported ec004 was not duplicated and the log data could not be retrieved. In the console ten codes of ec284 was found. A wet lab was performed and the ec284 was duplicated. The fluid level sensor was removed and replaced. After the replacement, f6 software was reinstalled to clear and remove the corrupted files. Mps console was then tested and passed the operational verification test. The root cause of the reported ec004 is unknown.
Patient Sequence No: 1, Text Type: N, H10


[139822987] A report was received regarding an alleged issue encountered during use of the console. The report states that during a case, the console displayed ec004. The user power cycled a few times where the message was observed after each power cycle. The console was turned off and on again and this resolved the issue. There were no reported patient complications resulting from the issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1649914-2019-00016
MDR Report Key8446288
Date Received2019-03-22
Date of Report2019-03-22
Date Mfgr Received2019-03-01
Date Added to Maude2019-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTOSAN ONOSODE
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal75002
Manufacturer Phone9723326338
Manufacturer G1QUEST MEDICAL, INC.
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal Code75002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameMPS 2 CONSOLE
Generic NameCPBP HEAT EXHANGER
Product CodeDTR
Date Received2019-03-22
Model Number5201260
Lot Number3017
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerQUEST MEDICAL, INC.
Manufacturer AddressONE ALLENTOWN PARKWAY ALLEN TX 75002 US 75002

Device Sequence Number: 1

Brand NameMPS 2 CONSOLE
Generic NameCPBP HEAT EXHANGER
Product CodeDTR
Date Received2019-03-22
Model Number5201260
Lot Number3017
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerQUEST MEDICAL, INC.
Manufacturer AddressONE ALLENTOWN PARKWAY ALLEN TX 75002 US 75002


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-22