RC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-03-25 for RC manufactured by Getinge Disinfection Ab.

Event Text Entries

[139686554] (b)(4). The issue is being investigated by manufacturing site.
Patient Sequence No: 1, Text Type: N, H10


[139686555] On (b)(6) 2019 we became aware of an issue with loading equipment outside machine- rc (return conveyor). As it was stated, the end pin of conveyor was stuck in "down" position resulting in situation when cart detached from it and fell down. The rc is a supplemental product to the washer disinfectors model number cm320 and 88-series registered as the medical device. There was no injury reported related to this issue. However having in mind similar cases and considering the worse case scenario we decided to report this issue based on the potential, as it might lead to an adverse event if it was to reoccur.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616031-2019-00010
MDR Report Key8447159
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-03-25
Date of Report2019-05-28
Date Facility Aware2018-03-18
Date Mfgr Received2019-03-18
Device Manufacturer Date2015-01-08
Date Added to Maude2019-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR DENNIS GENITO
Manufacturer Street45 BARBOUR POND DRIVE
Manufacturer CityWAYNE 07470
Manufacturer CountryUS
Manufacturer Postal07470
Manufacturer Phone9737097515
Manufacturer G1GETINGE DISINFECTION AB
Manufacturer StreetLJUNGADALSGATAN 11 KRONOBERGS LAN [SE-07]
Manufacturer CityVAXJO, 35115
Manufacturer CountrySW
Manufacturer Postal Code35115
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRC
Generic NameRC
Product CodePEC
Date Received2019-03-25
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGETINGE DISINFECTION AB
Manufacturer AddressLJUNGADALSGATAN 11 KRONOBERGS LAN [SE-07] VAXJO, 35115 SW 35115


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-25