[139701929] The field service engineer observed hydraulic fluid dripping from the hydraulic line that runs to the pressure gauge of the ort300 table. He replaced the or table hydraulic line that runs to the pressure gauge to resolve the issue.
Patient Sequence No: 1, Text Type: N, H10


[139701930] The customer service engineer was on-site to repair an ort300 table trolley. While inspecting the ort300 table, he observed hydraulic fluid dripping from the belly cover of the table while operating the hydraulic pump. There was no patient involved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010326005-2019-00004
MDR Report Key8447989
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-03-25
Date of Report2019-03-25
Date of Event2019-03-01
Date Mfgr Received2019-03-01
Device Manufacturer Date2014-12-09
Date Added to Maude2019-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD SPERLING
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal553434100
Manufacturer Phone7632036344
Manufacturer G1IMRIS-DEERFIELD IMAGING, INC.
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal Code553434100
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMRIS ORT300 OPERATING ROOM TABLE
Generic NameOPERATING ROOM TABLE
Product CodeKXJ
Date Received2019-03-25
Model NumberORT300
Catalog Number114093-000
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEERFIELD IMAGING, INC.
Manufacturer Address5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.