PRISMAFLO II PF2-265NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-26 for PRISMAFLO II PF2-265NA manufactured by Stihler Electronic Gmbh.

Event Text Entries

[139974252] The crrt circuit heater, which was described in the event description, is not a part of the prismaflex dialysis system manufactured by gambro (b)(4). This device is a blood warmer prismaflo ii manufactured by stihler electronic as an accessory to prismaflex dialysis system. It is intended to warm the blood return line during prismaflex crrt treatment. Because of the characteristics of this event the manufacturer judges the event fits in the recall for this product that addresses this potential failure. There was no patient injury or medical intervention associated with this event.
Patient Sequence No: 1, Text Type: N, H10


[139974253] It was reported that a crrt circuit heater malfunctioned and melted a patient's blood return line. Smoke/steam was present. Rn noticed that the circuit warmer hose was hot to touch with blisters on the surface. The crrt needed to be discontinued. Rn notified change immediately. There was no patient injury or medical intervention associated with this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617473-2019-00001
MDR Report Key8450977
Date Received2019-03-26
Date Mfgr Received2019-02-15
Date Added to Maude2019-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL SCHELLING
Manufacturer StreetJULIUS-HOELDER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal70597
Manufacturer G1STIHLER ELECTRONIC GMBH
Manufacturer StreetJULIUS-HOELDER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal Code70597
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-0761-2019
Event Type3
Type of Report0

Device Details

Brand NamePRISMAFLO II
Generic NameBLOOD WARMER
Product CodeKOC
Date Received2019-03-26
Model NumberPF2-265NA
Catalog NumberPF2-265NA
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTIHLER ELECTRONIC GMBH
Manufacturer AddressJULIUS-HOELDER-STRASSE 36 STUTTGART, 70597 GM 70597


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-26

© 2024 FDA.report