FACSARIA? FUSION 6B/3R/3V ACDU W/HOOD 656700G2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-28 for FACSARIA? FUSION 6B/3R/3V ACDU W/HOOD 656700G2 manufactured by Becton, Dickinson And Company, Bd Biosciences.

Event Text Entries

[140285046] The facsaria? Fusion is a product intended for research use only and this report is filed out of an abundance of caution due to serious injury. (b)(6). Device evaluated by mfr: a device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[140285047] It was reported that while using a facsaria? Fusion 6b/3r/3v acdu w/hood (a research use only, ruo, device) the lab technician attempted to change the fluid line from the 70% ethanol tank. The lab technician noted that the outside of the filter was wet. While the lab technician was holding the filter, the filter cap came loose and fluid from the 70% ethanol filter sprayed into the lab technician's face getting into their eyes. The lab technician washed their head and eyes with running water and wiped away the fluid. The lab technician sought medical attention form a physician in the department of internal medicine. The physician gave the lab technician a medical evaluation and assessed the condition of their eyes. The lab technician received further examination at the hospital's eye clinic which confirmed there was no abnormality or scarring in the patient's eyes.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916837-2019-00107
MDR Report Key8460491
Date Received2019-03-28
Date of Report2019-04-16
Date of Event2019-02-25
Date Mfgr Received2019-02-27
Device Manufacturer Date2015-04-29
Date Added to Maude2019-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Manufacturer Street2350 QUME DRIVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal Code95131
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFACSARIA? FUSION 6B/3R/3V ACDU W/HOOD
Generic NameFLOW CYTOMETER
Product CodeOYE
Date Received2019-03-28
Catalog Number656700G2
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Manufacturer Address2350 QUME DRIVE SAN JOSE CA 95131 US 95131


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-03-28