BARD? CRITICORE? MONITOR 000002N

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other,use report with the FDA on 2019-03-28 for BARD? CRITICORE? MONITOR 000002N manufactured by C.r. Bard, Inc. (covington) -1018233.

Event Text Entries

[140548485] The reported issue was confirmed. The root cause of the reported issue was isolated to a calibration issue. The temperature reading was calibrated. The monitor passed all tests and is functioning properly and ready for use. The device history record was reviewed and found nothing that could have caused or contributed to the reported event. The instructions for use were found adequate and state the following: "warning: the criticore monitor requires special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following tables. Portable and mobile rf communications equipment may affect the criticore monitor. Warning: the criticore monitor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the criticore monitor should be observed to verify normal operation in the configuration in which it will be used. Warning: do not immerse or submerge the monitor or turn it upside down when cleaning. Caution: use only heavy-duty alkaline d cell batteries. Do not use rechargeable batteries. Do not incinerate batteries. Recycle or dispose of them properly. Contact bard for disposal information. Caution: improper orientation of the batteries within the battery pack can potentially damage the criticore monitor. Caution: state and federal regulations govern the packaging necessary for return of medical product which may have been contaminated. Refer to local, state and federal regulations when packaging the criticore monitor for return. Caution: there are no user serviceable components inside the criticore monitor. The user should not attempt to repair the criticore monitor. To do so may void the warranty and could result in erroneous monitor readings. Caution: use of cables or sensors other than those specified for use with the criticore monitor, except those sold by bard for use as replacement part or repair components, may result in increased emissions or decreased immunity of the criticore monitor. Caution: the criticore monitor should be recycled properly per european union directive 2002/96/ec on waste electronic and electrical equipment, january 27, 2003. Do not dispose with ordinary municipal waste. "
Patient Sequence No: 1, Text Type: N, H10


[140548486] Per completion of the investigation, it was found that the observed measured temperature was reading low when the test temperature plug was used.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2019-01613
MDR Report Key8461640
Report SourceHEALTH PROFESSIONAL,OTHER,USE
Date Received2019-03-28
Date of Report2019-04-01
Date Mfgr Received2019-03-15
Device Manufacturer Date2012-12-01
Date Added to Maude2019-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYONIC ANDERSON
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1C.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal Code30014
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBARD? CRITICORE? MONITOR
Generic NameCRITICORE MONITOR
Product CodeEXS
Date Received2019-03-28
Returned To Mfg2019-02-12
Model Number000002N
Catalog Number000002N
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerC.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Address8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.