CLINICAL CHEMISTRY MAGNESIUM 03P68-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-03-29 for CLINICAL CHEMISTRY MAGNESIUM 03P68-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[140315385] This event was previously reported under manufacturer report number 1415939-2018-00168, with the incorrect manufacturing location. Any further information will be provided under this mdr report number. An investigation was performed by reviewing the complaint text, service history, tracking and trending metrics, and the current labeling for the magnesium package insert and the architect system operations manual. Review of complaint activity did not identify other complaints for the likely cause lot 99555un18. Tracking and trending for the magnesium assay determined there were no adverse or non-statistical trends. No nonconformances were identified related to this issue. A review of the product labeling concluded that the issue is sufficiently addressed. No systemic issue or deficiency of the architect magnesium assay was identified.
Patient Sequence No: 1, Text Type: N, H10


[140315386] The customer reported a falsely elevated architect magnesium initial result of 7. 18 mg/dl. Repeat result was 1. 88 mg/dl. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2019-00264
MDR Report Key8464199
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-03-29
Date of Report2019-03-29
Date of Event2018-11-25
Date Mfgr Received2019-03-05
Device Manufacturer Date2018-05-01
Date Added to Maude2019-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY MAGNESIUM
Generic NameMAGNESIUM
Product CodeJGJ
Date Received2019-03-29
Catalog Number03P68-21
Lot Number99555UN18
Device Expiration Date2019-03-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.