550 TXT TREATMENT TABLE 07346534

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2007-05-07 for 550 TXT TREATMENT TABLE 07346534 manufactured by .

Event Text Entries

[20345722] A product problem was reported with our treatment table. After treatment session was completed, the operator was lowering the table top by using a hand control device to allow patient dismount. The table abruptly stopped and the operator attempted to lower the table again by using the hand control device. The table top suddenly dropped approximately 10cm without producing an interlock during non-treatment state. Root cause on the sudden vertical drop has not been determine and is currently under investigation. No injury was reported as a result of this incident. It's important to note, by design system will trigger an interlock if table drops beyond specification during treatment. In addition, the 10cm drop was not measured with a calibrated device or based on captured data from the system, and only an approximate distance that was reported from the site.
Patient Sequence No: 1, Text Type: D, B5


[20452177] Investigation is still in process; therefore, no conclusion can be made at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2910081-2007-00015
MDR Report Key846456
Report Source01,05,06
Date Received2007-05-07
Date of Report2007-04-16
Date of Event2007-04-14
Date Mfgr Received2007-04-16
Device Manufacturer Date2005-03-01
Date Added to Maude2007-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAN TAKEMOTO
Manufacturer Street4040 NELSON AVE.
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal94520
Manufacturer Phone9256028192
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name550 TXT TREATMENT TABLE
Generic NameCOUCH, RADIATION, THERAPY, POWERED
Product CodeJAI
Date Received2007-05-07
Model Number07346534
ID Number3845
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key833873


Patients

Patient NumberTreatmentOutcomeDate
10 2007-05-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.