STERRAD? SEALSURE CI TAPE 14202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-03-29 for STERRAD? SEALSURE CI TAPE 14202 manufactured by Advanced Sterilization Products.

Event Text Entries

[140325852] Asp investigation summary: the investigation included a review of the device history record, trending analysis by lot number, system risk analysis (sra) and supplier product investigation. The device history record was reviewed and process specifications for product release were met. No issues were observed that would contribute to the complaint. Trending analysis by lot number was reviewed from 2/6/2018 to 8/6/2018 and no significant trend was observed. The sra indicates the risk associated with exposure to toxic or corrosive material is "low. ". One sterrad? Sealsure ci tape retains of the same lot was subject to functional investigation by the supplier. A sterilization test was performed and the color of the ci tape changed normally from red to yellow. The assignable cause of the issue could not be confirmed; however, a likely cause could be due to the customer's load contents as subsequent use of the sealsure chemical indicator tape provided acceptable results according to the customer. Also, the supplier could not replicate the issue with the retain sample. The issue will continue to be tracked and trended. (b)(6). Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[140325853] A customer initially reported the sterrad? Sealsure chemical indicator tape did not change color correctly after a completed sterrad? 100s cycle. At the time of the report, it was unknown whether or not the affected load was released for use on patients prior to reprocessing. Upon follow-up with the customer on (b)(6) 2019, it was determined the load was released for use without reprocessing first. There is no report of infection, injury or harm to patient(s) associated with this issue. Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured. Therefore, as a matter of policy asp had decided to report all incidents of sterrad? Sealsure chemical indicator tape not changing color correctly when the load is released for use without reprocessing first.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2019-00846
MDR Report Key8464847
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-03-29
Date of Report2019-03-07
Date of Event2018-08-06
Date Mfgr Received2019-03-29
Device Manufacturer Date2017-12-14
Date Added to Maude2019-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. GABRIEL ALFAGEME
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone949789-868
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERRAD? SEALSURE CI TAPE
Generic NameINDICATOR, CHEMICAL
Product CodeFRC
Date Received2019-03-29
Catalog Number14202
Lot Number34817
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.