ELECSYS T4 ASSAY 07027885190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-01 for ELECSYS T4 ASSAY 07027885190 manufactured by Roche Diagnostics.

Event Text Entries

[140614671] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[140614672] The initial reporter complained of high thyroid results for 1 patient tested on a cobas e 801 module compared to the siemens centaur method. Based on the data provided, discrepant results were identified for elecsys t4 (t4), elecsys t3 (t3), elecsys ft3 iii (ft3 iii) and elecsys ft4 ii (ft4 ii). This medwatch will cover t4. Refer to medwatch with patient identifier (b)(6) for information on the t3 results, medwatch with patient identifier (b)(6) for information on the ft3 iii results and medwatch with patient identifier (b)(6) for information on the ft4 ii results. The results from the e801 module were reported outside of the laboratory. The results from the siemens method were believed to be correct. There was no allegation that an adverse event occurred. The e801 module serial number was (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01276
MDR Report Key8469934
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-01
Date of Report2019-06-21
Date of Event2019-03-07
Date Mfgr Received2019-03-14
Date Added to Maude2019-04-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS T4 ASSAY
Generic NameRADIOIMMUNOASSAY, TOTAL THYROXINE
Product CodeCDX
Date Received2019-04-01
Model NumberNA
Catalog Number07027885190
Lot Number35935902
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.