XPRESS MINIMALLY INVASIVE PEDICEL SCREW SYSTEM T073-1000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-04-01 for XPRESS MINIMALLY INVASIVE PEDICEL SCREW SYSTEM T073-1000 manufactured by X-spine Systems, Inc..

Event Text Entries

[140616434] The complainant reported that while inserting a rod, the distal tip of an inserter broke. The surgeon was able to successfully finish the surgery. There were no known patient complications. The instrument has not been received for complaint investigation. Pictures of the device depicted one of the distal hooks of the inserter broken from the instrument. A dhr review was performed. The instrument met all required specifications prior to being released to distributable inventory on 06/20/2016. A lateral force applied to the inserter while placing a rod into a pedicle screw may have caused the damage to the inserter. Application of lateral force to the inserter concentrates force to the interface between the inserter and rod.
Patient Sequence No: 1, Text Type: N, H10


[140616435] The complainant reported that while inserting a rod, the distal tip of an inserter broke. The surgeon was able to successfully finish the surgery. There was no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005031160-2019-00019
MDR Report Key8471542
Report SourceDISTRIBUTOR
Date Received2019-04-01
Date of Report2019-04-01
Date of Event2019-03-02
Date Mfgr Received2019-03-04
Device Manufacturer Date2016-06-20
Date Added to Maude2019-04-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY MING
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal59714
Manufacturer Phone4063880480
Manufacturer G1X-SPINE SYSTEMS, INC.
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal Code59714
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXPRESS MINIMALLY INVASIVE PEDICEL SCREW SYSTEM
Generic NamePEDICLE SCREW SPINAL SYSTEM
Product CodeHTC
Date Received2019-04-01
Model NumberT073-1000
Lot NumberXS157167A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerX-SPINE SYSTEMS, INC.
Manufacturer Address664 CRUISER LANE BELGRADE MT 59714 US 59714


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.