MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-02 for ELECSYS VITAMIN D TOTAL II 07464215190 manufactured by Roche Diagnostics.
[140621744]
This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[140621745]
The initial reporter stated that they noticed an issue with one reagent kit of the elecsys vitamin d total gen. 2 assay on the cobas 8000 e 602 module. An unspecified number of patient samples resulted with values of > 250 nmol/l and when repeated on another analyzer, results were within the technical limit of the assay. The customer provided data for a total of 36 patient samples and of these, one had an erroneous vitamin d result. The erroneous result was not reported outside of the laboratory. The sample initially resulted with a value of 88. 89 nmol/l when measured with the problematic kit. When tested on the other analyzer, the vitamin d result was 48. 98 nmol/l. No adverse events were alleged to have occurred with the patient. The e 602 analyzer serial number is (b)(4).
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2019-01317 |
MDR Report Key | 8472957 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2019-04-02 |
Date of Report | 2019-09-23 |
Date of Event | 2019-02-28 |
Date Mfgr Received | 2019-03-15 |
Date Added to Maude | 2019-04-02 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ELECSYS VITAMIN D TOTAL II |
Generic Name | VITAMIN D TEST SYSTEM |
Product Code | MRG |
Date Received | 2019-04-02 |
Model Number | NA |
Catalog Number | 07464215190 |
Lot Number | ASKU |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-02 |