TOTAL BILIRUBIN 06L45-41

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-04-02 for TOTAL BILIRUBIN 06L45-41 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[140750173] Patient identifier: multiple = (b)(6). There is no further patient information provided due to privacy issues. An evaluation is in process. A final report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[140750174] The customer reported falsely elevated total bilirubin results on two patients generated on the architect analyzer. The results provided were: sid (b)(6) initial = 152. 2 / repeat 6. 0umol/l; sid (b)(6) initial = 110. 9 / repeat =13. 3umol//l. The medic questioned the results prior to clinical decisions being made. There was no reported impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2019-00280
MDR Report Key8475150
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-04-02
Date of Report2019-04-29
Date of Event2019-03-09
Date Mfgr Received2019-04-10
Date Added to Maude2019-04-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOTAL BILIRUBIN
Generic NameTOTAL BILIRUBIN REAGENT
Product CodeCIG
Date Received2019-04-02
Catalog Number06L45-41
Lot Number54131UQ11
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.