MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2019-04-03 for SURGICAL CLIPPER, CHARGING BASE CAH4414 manufactured by Jiangsu Better Electric.
[140950561]
Based on the supplier investigation the device history record review did not indicate any exception that could lead to the reported incident. No sample was available for investigation, only a photo was provided. The supplier checked the device master record and there have been no changes since product launch. All units go through 100% in-process system burn-in within 48 hours to screen out workmanship problems and premature defects and no abnormality was found in the record. From the investigation, the root cause could not be determined. There is no action taken at this time, but we will continue to monitor the trend of this type of incident.
Patient Sequence No: 1, Text Type: N, H10
[140950562]
A burnt odor was coming from surgical clipper charging base so the customer stopped using it. There was no injury.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1423537-2019-00301 |
MDR Report Key | 8480358 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2019-04-03 |
Date of Report | 2019-04-03 |
Date of Event | 2019-03-13 |
Date Facility Aware | 2019-03-14 |
Date Added to Maude | 2019-04-03 |
Event Key | 0 |
Report Source Code | Distributor report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PATRICIA TUCKER |
Manufacturer Street | 3651 BIRCHWOOD DRIVE |
Manufacturer City | WAUKEGAN IL 60085 |
Manufacturer Country | US |
Manufacturer Postal | 60085 |
Manufacturer Phone | 8478874151 |
Manufacturer G1 | JIANGSU BETTER ELECTRIC |
Manufacturer Street | 11 HUI HAI RD. BIN HAI ZONE, Q |
Manufacturer City | JIANG SU 226200 |
Manufacturer Country | CH |
Manufacturer Postal Code | 226200 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SURGICAL CLIPPER, CHARGING BASE |
Generic Name | RAZOR, SURGICAL |
Product Code | LWK |
Date Received | 2019-04-03 |
Model Number | CAH4414 |
Catalog Number | CAH4414 |
Lot Number | 1424-1377 |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | JIANGSU BETTER ELECTRIC |
Manufacturer Address | 11 HUI HAI RD. BIN HAI ZONE, Q JIANG SU 226200 CH 226200 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-03 |