ALBP ALBUMIN BCP ALB BCP 05975573190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-04 for ALBP ALBUMIN BCP ALB BCP 05975573190 manufactured by Roche Diagnostics.

Event Text Entries

[141107763] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[141107764] The initial reporter complained of discrepant results for 3 patient samples tested for albp albumin bcp (albp) on a cobas 8000 c 702 module. Patient 1 initial albp result was 68 g/l. The repeat result was 43 g/l. Patient 2 initial albp result was 58 g/l. The repeat result was 40 g/l. Patient 3 initial albp result was 81 g/l. The repeat result was 43 g/l. No erroneous results were reported outside of the laboratory. There was no allegation that an adverse event occurred. The c702 module serial number was (b)(4). The field service engineer (fse) visited the customer site and adjusted the sample probe and rinse unit. One of the sample probes was replaced by the customer. The customer has not provided any additional information for investigation. The investigation did not identify a product problem. The cause of the event could not be determined.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01365
MDR Report Key8483242
Date Received2019-04-04
Date of Report2019-04-04
Date of Event2018-10-04
Date Mfgr Received2019-03-12
Date Added to Maude2019-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameALBP ALBUMIN BCP
Generic NameBROMCRESOL PURPLE DYE-BINDING, ALBUMIN
Product CodeCJW
Date Received2019-04-04
Model NumberALB BCP
Catalog Number05975573190
Lot Number35651401
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.