[140935649] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[140935650] According to the reporter, when they connected the catheter, they received an error "catheter not supported/catheter not connected" and the black piece that locks into the syringe was not moving to the correct position to properly fit the catheter. It took three catheters to get the problem done. The machine was turned off and on and the catheter was disconnected and reconnected, but it did not fix the problem. Technical support confirmed that the catheter connector was inserted with the green dots aligned. There was no issue with the syringe plunger and the catheter display did not have a real-time clock battery depleted/cmos error. They had an issue with archiving the data, but not the current problem, and the catheter was not difficult to remove from the patient. There was no patient and user harm, but they required an additional anesthesia.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006897778-2019-00005
MDR Report Key8483521
Report SourceUSER FACILITY
Date Received2019-04-04
Date of Report2019-04-04
Date of Event2019-01-11
Date Mfgr Received2019-03-14
Date Added to Maude2019-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAMY BEEMAN
Manufacturer Street161 CHESHIRE LANE, SUITE 100
Manufacturer CityPLYMOUTH MN 55441
Manufacturer CountryUS
Manufacturer Postal55441
Manufacturer Phone7632104064
Manufacturer G1CROSPON LTD
Manufacturer StreetGALWAY BUSINESS PARK, DANGAN
Manufacturer CityGALWAY,FL H91P2DK
Manufacturer Postal CodeH91P2DK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOFLIP
Generic NameSYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)
Product CodeFFX
Date Received2019-04-04
Model NumberEF-322N
Catalog NumberEF-322N
Lot Number322N180718
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCROSPON LTD
Manufacturer AddressGALWAY BUSINESS PARK, DANGAN GALWAY,FL H91P2DK H91P2DK


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-04