GORE VIATORR TIPS ENDOPROSTHESIS WITH CONTROLLED EXPANSION PTB8108275

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-05 for GORE VIATORR TIPS ENDOPROSTHESIS WITH CONTROLLED EXPANSION PTB8108275 manufactured by W.l. Gore & Associates.

Event Text Entries

[141168188] The patient weight or date of birth were not available. The imaging evaluation was able to confirm the physician? S event description that balloon thrombectomy was being performed. No images were provided of the device after it had been dislodged to the right atrium, so the distance that the device was dislodged was unable to be confirmed.
Patient Sequence No: 1, Text Type: N, H10


[141168189] It was reported the physician deployed a gore? Viatorr? Tips endoprosthesis. The patient had preexisting portal vein thrombosis. One day later, the physician was performing a balloon thrombectomy. During one of the passes, the viatorr? Device was dislodged to the right atrium. The physician was able to pull the expanded device out through the internal jugular, which enlarged the access site. A 16 fr sheath with a suture around it was used to achieve hemostasis. A vascular surgeon closed the access site once the procedure was finished. The patient was doing well following the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007284313-2019-00098
MDR Report Key8489480
Date Received2019-04-05
Date of Report2019-04-11
Date of Event2018-12-15
Device Manufacturer Date2018-09-10
Date Added to Maude2019-04-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAN KITTERMAN
Manufacturer Street1500 N. 4TH STREET
Manufacturer CityAZ
Manufacturer Phone9285263030
Manufacturer G1MEDICAL PHOENIX 1 B/P
Manufacturer Street32360 N. NORTH VALLEY PARKWAY
Manufacturer CityPHOENIX AZ 85085
Manufacturer CountryUS
Manufacturer Postal Code85085
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameGORE VIATORR TIPS ENDOPROSTHESIS WITH CONTROLLED EXPANSION
Generic NameSHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Product CodeMIR
Date Received2019-04-05
Catalog NumberPTB8108275
Lot Number18570143
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerW.L. GORE & ASSOCIATES
Manufacturer AddressFLAGSTAFF AZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-05

© 2026 FDA.report