MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-10 for MDT-SLITTER manufactured by Medtronic, Inc..
[141911366]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[141911367]
It was reported that during the implant procedure the physician was slitting the catheter and the slitter slipped off. It was also reported the physician had difficulty affixing to start slitting again and this process the right ventricular (rv) lead insulation was damaged. It was noted the lead numbers were stable despite the insulation breach. The lead was removed and replaced. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2182208-2019-00674 |
MDR Report Key | 8498655 |
Date Received | 2019-04-10 |
Date of Report | 2019-04-10 |
Date of Event | 2019-03-14 |
Date Mfgr Received | 2019-03-14 |
Date Added to Maude | 2019-04-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LISA ROBERTSON |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635262723 |
Manufacturer G1 | MEDTRONIC, INC. |
Manufacturer Street | 8200 CORAL SEA STREET NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal Code | 55112 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Generic Name | MAGNET, TEST, PACEMAKER |
Product Code | DTG |
Date Received | 2019-04-10 |
Model Number | MDT-SLITTER |
Catalog Number | MDT-SLITTER |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC, INC. |
Manufacturer Address | 8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-10 |