ESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR 81-040412

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-04-10 for ESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR 81-040412 manufactured by Deroyal Cientifica De Latinoamerica S.r.l..

Event Text Entries

[146657064] Root cause: a root cause could not be determined. The defective sample was not returned for evaluation, and no issues were identified during a review of quality records. Corrective action: due to the root cause determination, a corrective action has not been taken. Investigative summary: an internal complaint (call (b)(4)) was received indicating an esophageal stethoscope with temperature sensor (part 81-040412, lot 48181841) stopped detecting temperature during use, and a skin probe had to be pulled to continue monitoring the patient's temperature. The defective sample was unavailable for return. Therefore, a functional inspection could not be performed. The initial report indicated representative samples from the same lot were available for return. However, as of the date of this report, those samples have not been returned to the manufacturer for evaluation. The device history record (dhr) was reviewed for the reported lot number and no issues were reported during the manufacturing or inspection of the product. The bill of material for the work order involved was reviewed to identify the sub-assembly wire set used to manufacture the lot. The functional results for this sub-assembly (lot 48242691) were reviewed and no discrepancies were reported during the quality inspection. A functional inspection of product in process was made. One-hundred and twenty-five samples from work order 48229429 were inspected and all units were found within tolerance. The 2017-2019 complaint log was reviewed for similar complaints. Two complaints ((b)(4)) were identified for an esophageal stethoscope displaying inconsistent temperature readings. At this time, these complaints have been closed and verified. The investigation is complete at this time. If new and critical information is received, this report will be updated.
Patient Sequence No: 1, Text Type: N, H10


[146657065] During a dental case, the esophageal stethoscope with temperature sensor stopped reading the patient's temperature. A skin probe was pulled to the patient's temperature could continue being monitored.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9613793-2019-00002
MDR Report Key8500822
Report SourceUSER FACILITY
Date Received2019-04-10
Date of Report2019-04-10
Date of Event2019-03-07
Date Mfgr Received2019-03-11
Date Added to Maude2019-04-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSARAH BENNETT
Manufacturer Street200 DEBUSK LANE
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8653626112
Manufacturer G1DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L.
Manufacturer StreetGLOBAL PARK BOX 180-3006, 602 PARKWAY
Manufacturer CityLA AURORA, HEREDIA, HEREDIA 146
Manufacturer CountryCS
Manufacturer Postal Code146
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR
Generic NameSTETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS
Product CodeBZT
Date Received2019-04-10
Model Number81-040412
Lot Number48181841
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEROYAL CIENTIFICA DE LATINOAMERICA S.R.L.
Manufacturer AddressGLOBAL PARK BOX 180-3006, 602 PARKWAY LA AURORA, HEREDIA, HEREDIA 146 CS 146


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.