ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR 04388780190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-11 for ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR 04388780190 manufactured by Roche Diagnostics.

Event Text Entries

[141588060] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[141588061] The initial reporter complained of discrepant high results for 10 patient samples tested for elecsys anti-tshr immunoassay (anti-tshr) on a cobas 8000 e 602 module. No erroneous results were reported outside of the laboratory. The repeat results were believed to be correct. There was no allegation that an adverse event occurred. The e602 module serial number was (b)(4). The discrepant results were all from the same reagent pack and occurred within a short time period. Qc results were acceptable. The customer was not having problems with other assays run on the instrument. Calibration and qc were acceptable. Based on the calibration and qc data, a general reagent issue can be excluded. The issue may have been due to the presence of foam/bubbles on the surface of the reagent, however, the investigation was unable to determine the specific root cause of the event. The investigation did not identify a product problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01427
MDR Report Key8503037
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-11
Date of Report2019-04-11
Date of Event2019-03-16
Date Mfgr Received2019-03-26
Date Added to Maude2019-04-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameANTI-TSHR, ANTIBODIES TO TSH RECEPTOR
Generic NameANTI-TSHR IMMUNOASSAY
Product CodeJZO
Date Received2019-04-11
Model NumberNA
Catalog Number04388780190
Lot Number33584301
Device Expiration Date2019-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.