SHEARGUARD CHEST PAD COVERS 5844-12

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2019-04-12 for SHEARGUARD CHEST PAD COVERS 5844-12 manufactured by Mizuho Orthopedic Systems, Inc..

Event Text Entries

[141763660] Unable to determine if the patient had comorbidities that could have contributed to the problem. Advised the distributor to discuss the owner's manual positioning instructions with the customer. This problem was initially reported as related to the proaxis table. Investigation found it might also be related to the two single-use pad covers, so two additional reports are being filed showing the possible association of the injury with those two devices.
Patient Sequence No: 1, Text Type: N, H10


[141763661] Patient was positioned prone for posterior approach surgery utilizing the standard proaxis spine assembly (open frame/spars) configured with torso trolley, arm boards, chest/pelvic bolsters and leg support. All items were covered prior to the procedure with the recommended osi shear protect consumables to minimize risk of skin injury. Postoperatively, the client presented with pressure related sores under areas of contact around sternum and asis. It was noted that bmi and length of case (up to 13 hours) could have been contributing factors.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2921578-2019-00015
MDR Report Key8509085
Report SourceDISTRIBUTOR,FOREIGN
Date Received2019-04-12
Date of Report2019-08-28
Date of Event2019-01-08
Date Mfgr Received2019-04-02
Date Added to Maude2019-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE LEBLANC
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal945871234
Manufacturer Phone5104291500
Manufacturer G1MIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal Code945871234
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSHEARGUARD CHEST PAD COVERS
Generic NameCOVERS
Product CodeKME
Date Received2019-04-12
Model Number5844-12
Catalog Number5844-12
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Address30031 AHERN AVENUE UNION CITY CA 945871234 US 945871234


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.