PROTEGRA *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-05-11 for PROTEGRA * manufactured by Mavig Gmbh.

Event Text Entries

[21709422] There have been three recent incidents of cracks occurring in the articulating arms manufactured by mavig used to support power injectors and lead shields. Two of these cracks have resulted in the arm to shear off completely and fall to the ground. The other arm did not fall; although a significant crack in the arm casing has been noted. All mavig arms have since been evaluated and those with defects or cracks have been removed from service entirely.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number851066
MDR Report Key851066
Date Received2007-05-11
Date of Report2007-05-11
Date of Event2007-04-10
Report Date2007-05-11
Date Reported to FDA2007-05-11
Date Added to Maude2007-05-17
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NamePROTEGRA
Generic NameARM, ARTICULATING, EQUIPMENT SUPPORT
Product CodeKPY
Date Received2007-05-11
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age6 YR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key837813
ManufacturerMAVIG GMBH
Manufacturer AddressCARE OF: TI-BA ENTERPRISES, IN 25 HYTEC CIRCLE ROCHESTER NY 14606 US

Device Sequence Number: 2

Brand NamePROTEGRA
Generic NameARM, ARTICULATING, EQUIPMENT SUPPORT
Product CodeKPY
Date Received2007-05-11
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age5 YR
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key837814
ManufacturerMAVIG GMBH
Manufacturer AddressCARE OF: TI-BA ENTERPRISES, IN 25 HYTEC CIRCLE ROCHESTER NY 14606 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-05-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.