PAL-650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-12 for PAL-650 manufactured by Microaire Surgical Instruments.

Event Text Entries

[145618953] On april 5, 2019, pal-650/serial#: (b)(4), was subjected to internal acceptance testing. The device passed all testing and it was not possible to confirm that there were any issues with the device's functionality. A follow-up phone call between microaire formally designated complaint handling unit and a surgical technician at the user facility took place and it was reported that at times, those in charge of cleaning/sterilization are "in a rush" and do not allow their pal-650 to dry adequately. The surgical technician was instructed to follow the instructions for use associated with the device, and that not allowing adequate dry time can cause intermittent failures of the device. It is likely that the reported issue was caused by the device not being allowed to dry adequately. A review of complaint files has revealed that pal-650 is performing as expected, with regard to this failure mode.
Patient Sequence No: 1, Text Type: N, H10


[145618954] On (b)(6) 2019, it was reported that during a liposuction procedure, a pal-650 handpiece would not function. The doctor had to perform the procedure manually until a second pal-650 device was sterilized. This delayed the surgery approximately 60 minutes. There was no patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020601-2019-00003
MDR Report Key8511710
Date Received2019-04-12
Date of Report2019-04-12
Date of Event2019-03-15
Date Mfgr Received2019-03-15
Date Added to Maude2019-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN AMELL
Manufacturer Street3590 GRAND FORKS BLVD.
Manufacturer CityCHARLOTTESVILLE VA 22902
Manufacturer CountryUS
Manufacturer Postal22902
Manufacturer Phone4349758394
Manufacturer G1MICROAIRE SURGICAL INSTRUMENTS
Manufacturer Street3590 GRAND FORKS BLVD.
Manufacturer CityCHARLOTTESVILLE VA 22911
Manufacturer CountryUS
Manufacturer Postal Code22911
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePAL-650
Generic NamePOWER ASSISTED LIPOPLASTY
Product CodeMUU
Date Received2019-04-12
Returned To Mfg2019-04-04
Model NumberPAL-650
Catalog NumberPAL-650
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROAIRE SURGICAL INSTRUMENTS
Manufacturer Address3590 GRAND FORKS BLVD. CHARLOTTESVILLE VA 22911 US 22911