[141912610] A zeiss field service engineer (fse) performed an on-site inspection of the microscope. The fse determined that the binoculars fell off as a result of a loose set screw, which had not been sufficiently tightened by the user, and not as a result of a device malfunction. It was concluded that the user did not check and firmly tighten the set screw before using the microscope. The user manual gives clear instruction to make sure that all accessories are securely mounted. The user has been reinformed about the importance of performing safety and function checks before every use of the microscope.
Patient Sequence No: 1, Text Type: N, H10


[141912611] The binocular of the opmi pico microscope mounted on a s100 floor stand detached and fell on a doctor. It was reported that the doctor sustained a bruise on her shoulder. No additional information could be obtained from the site regarding the extent of the injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2019-00003
MDR Report Key8511738
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-04-12
Date of Report2019-03-13
Date of Event2019-03-13
Date Mfgr Received2019-03-13
Device Manufacturer Date2011-04-11
Date Added to Maude2019-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. VERNON BROWN
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574616
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 FLOORSTAND
Generic NameOPMI PICO
Product CodeFSO
Date Received2019-04-12
Model NumberN/A
Catalog Number000000-1095-102
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.