8MM SINGLE USE OATS SET AR-1981-08S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2007-05-11 for 8MM SINGLE USE OATS SET AR-1981-08S manufactured by Arthrex, Inc..

Event Text Entries

[646018] It was reported that the recipient piece broke during the harvesting of the recipient socket, 20mm from the end, just above the window. The procedure was normal. The recipient has been used multiple times. The recipient broke during the third time. The broken piece remains in the pt. This device is used for treatment.
Patient Sequence No: 1, Text Type: D, B5


[7965468] Eval confirmed the recipient harvester broke at the 20mm laser marking. Material verification was performed and found within specifications. This product is a single use osteochondral autograft transfer system. The original info provided indicates the device was used multiple times. However, no clarification could be obtained from the customer, as if the device was used several times during the same surgery or during different surgeries. This is the first complaint of this type for this product. A definitive conclusion could not be determined from the info provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1220246-2007-00078
MDR Report Key851331
Report Source01,08
Date Received2007-05-11
Date of Report2007-05-10
Date of Event2007-01-01
Date Mfgr Received2007-04-16
Device Manufacturer Date2006-09-01
Date Added to Maude2007-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactIVETTE GALMEZ, SR. ANALYST
Manufacturer Street1370 CREEKSIDE BLVD.
Manufacturer CityNAPLES FL 341081945
Manufacturer CountryUS
Manufacturer Postal341081945
Manufacturer Phone8009337001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name8MM SINGLE USE OATS SET
Generic NameHARVESTING TOOLS
Product CodeGFI
Date Received2007-05-11
Returned To Mfg2007-04-30
Model NumberNA
Catalog NumberAR-1981-08S
Lot Number06124
ID NumberNA
Device Expiration Date2011-09-01
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key838394
ManufacturerARTHREX, INC.
Manufacturer AddressNAPLES FL US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-05-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.