ELECSYS VITAMIN D TOTAL II 07464215190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-15 for ELECSYS VITAMIN D TOTAL II 07464215190 manufactured by Roche Diagnostics.

Event Text Entries

[141939556] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[141939637] The initial reporter questioned high results for 2 patient samples and qc results tested for elecsys vitamin d total ii (vitamin d total ii) on a cobas e 411 immunoassay analyzer. The reagent kit was received by the customer on 15-mar-2019; the customer did not perform new calibration for this particular reagent kit. The system used the previous calibration as is designed. The qc results had been within the acceptable range, but the qc and patient results were now higher than expected at > 100 ng/ml. The customer repeated these tests by diluting the samples 1:3 but the results were still > 100 ng/ml. The customer cleaned the sample/reagent probe and repeated qc and patient results. Discrepant patient results were received: patient 1 initial result was > 100 ng/ml. The repeat result was 20. 21 ng/ml. Patient 2 initial result was > 100 ng/ml. The repeat result was 30. 63 ng/ml. The customer also continued to have issue with qc results. It is not known if the high patient results were reported outside of the laboratory. There was no allegation that an adverse event occurred. Roche diagnostics has issued an urgent medical device correction for this issue, entitled "elecsys? Vitamin d total ii assay? Non-reproducible false high results. " this correction has been reported to fda. During the implementation of the elecsys vitamin d total ii assay on the modular analytics e 170 module, and cobas e 601 and 602 systems, there were reports of non-reproducible, false high results. These customer observations were made in comparisons of duplicate measurements during assay validation of elecsys vitamin d total ii assay or in method comparisons with elecsys vitamin d assay (i. E. , during conversions), where the falsely elevated discrepant value did not fit the expected result. The observed elevated results reported with the elecsys vitamin d ii assay were not confirmed upon repeat testing of the affected samples. "the observed findings: the first result is elevated, either above the upper end of the measuring range (>100 ng/ml or >250 nmol/ml), or within the measuring range; repeated results are significantly lower. Rare cases have been reported on the cobas e 411 analyzer and the cobas e 801 module. Roche is conducting investigations into the reported issue and has determined that the elecsys? Vitamin d total ii assay is strongly affected by pre-analytical errors. The investigations have reproduced these findings when requirements for pre-analytical sample handling have not been met for plasma samples. Further investigations into the reported issue are ongoing. As a result, the pre-analytical sample quality and compliance to the tube manufacturer? S specifications is very important to assure a high quality sample in order to minimize the risk of occurrence. " investigations are ongoing to determine the root cause of this issue. A workaround has been provided to customers.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01437
MDR Report Key8513729
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-15
Date of Report2019-07-09
Date of Event2019-03-26
Date Mfgr Received2019-03-26
Date Added to Maude2019-04-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameELECSYS VITAMIN D TOTAL II
Generic NameVITAMIN D TEST SYSTEM
Product CodeMRG
Date Received2019-04-15
Model NumberNA
Catalog Number07464215190
Lot Number366696
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.