S-12 SURGICAL HANDPIECE 30061000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-16 for S-12 SURGICAL HANDPIECE 30061000 manufactured by W&h Dentalwerk Buermoos Gmbh.

Event Text Entries

[142037547] During a dental surgica treatment the instrument got hot and it ended up in burning the patient's lip (2nd degree). The injury is reversible and the patient is under monitoring. After evaluation of the suspect device it came out that the adverse event occured due to wrong lubrication of the instrument.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681479-2019-00002
MDR Report Key8517118
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-16
Date of Report2019-03-18
Date of Event2019-03-15
Date Mfgr Received2019-03-18
Device Manufacturer Date2018-09-27
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANJA LINDNER
Manufacturer StreetIGNAZ-GLASER-STRASSE 53
Manufacturer CityBUERMOOS, SALZBURG 5111
Manufacturer CountryAU
Manufacturer Postal5111
Manufacturer G1W&H DENTALWERK BUERMOOS GMBH
Manufacturer StreetIGNAZ-GLASER-STRASSE 53
Manufacturer CityBUERMOOS, SALZBURG 5111
Manufacturer CountryAU
Manufacturer Postal Code5111
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameS-12
Generic NameSURGICAL HANDPIECE
Product CodeKMW
Date Received2019-04-16
Model NumberSURGICAL HANDPIECE
Catalog Number30061000
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerW&H DENTALWERK BUERMOOS GMBH
Manufacturer AddressIGNAZ-GLASER-STRASSE 53 BUERMOOS, SALZBURG 5111 AU 5111


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2019-04-16

© 2026 FDA.report