N ANTISERUM TO HUMAN ALBUMIN 10446283

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-16 for N ANTISERUM TO HUMAN ALBUMIN 10446283 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[142665873] The customer contacted a siemens customer care center (ccc) and reported that a discordant, falsely elevated albumin result was obtained on a patient sample on the bn prospec system. Siemens analyzed the provided customer save data file. Siemens' analysis of the kinetic data from the complaint shows an increase in reactivity for the high sample result. Siemens is investigating the issue.
Patient Sequence No: 1, Text Type: N, H10


[142665874] A discordant, falsely elevated albumin result was obtained on a (b)(6) female patient sample on the bn prospec system using n antiserum to human albumin lot 153939. The albumin was used for calculating csf/albumin serum ratio. The initial discordant, falsely elevated result was not reported to the physician(s). The sample was repeated on the same reagent and system and resulted in a lower value. There is no indication of a corrected report being issued to the physician(s). There is no known report of patient intervention or adverse health consequences due to the discordant, falsely elevated albumin results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2019-00040
MDR Report Key8517881
Date Received2019-04-16
Date of Report2019-06-27
Date of Event2019-02-13
Date Mfgr Received2019-05-28
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BENLEY SAINT-JEAN
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10951
Manufacturer CountryUS
Manufacturer Postal10951
Manufacturer Phone9145242398
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL-VON-BEHRING-STR. 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Previous Use Code3
Removal Correction Number9610806-2019/06/28-001-C
Event Type3
Type of Report0

Device Details

Brand NameN ANTISERUM TO HUMAN ALBUMIN
Generic NameN ANTISERUM TO HUMAN ALBUMIN
Product CodeDCF
Date Received2019-04-16
Model NumberN ANTISERUM TO HUMAN ALBUMIN
Catalog Number10446283
Lot Number153939E
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL-VON-BEHRING- STR. 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.