CRYOMACS FREEZING BAGS 50 N/A 200-074-400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-16 for CRYOMACS FREEZING BAGS 50 N/A 200-074-400 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[146090813] The reason for the breakage was a clear application failure by the customer. The bags were filled with a higher volume than clearly indicated in the product information sheet. An overfilling results in mechanical stress to the bag that can easily lead to a breakage. This was the first complaint with this failure description for this lot. This one complaint refers to 4 bags. Complaint rate (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[146090814] The customer complains four cracked freezing bags 50 of lot 6170829012. The freezing bags were cracked at the fold line from bag and label area. The loaded volume was over 30 ml while the product information sheet states a recommended filling volume of 10-20 ml. No overwrap bag was used. Due to the loss of material a second treatment, that might be necessary under certain circumstances, would not be possible.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2019-00002
MDR Report Key8518652
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-16
Date of Report2019-04-16
Date of Event2019-03-25
Date Facility Aware2019-03-26
Date Mfgr Received2019-03-25
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. BETTINA-JUDITH H
Manufacturer StreetFRIEDRICH-EBERT-STR. 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal51429
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCRYOMACS FREEZING BAGS 50
Generic NameCRYOMACS FREEZING BAGS 50
Product CodeKSR
Date Received2019-04-16
Model NumberN/A
Catalog Number200-074-400
Lot Number6170829012
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STR. 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-16

© 2026 FDA.report