ELECSYS VITAMIN D TOTAL II 07464215190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-16 for ELECSYS VITAMIN D TOTAL II 07464215190 manufactured by Roche Diagnostics.

Event Text Entries

[142218130] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[142218131] The initial reporter complained of high results not corresponding to the clinical picture for 1 patient sample tested for elecsys vitamin d total ii (vitamin d ii) on a cobas 6000 e 601 module. The initial vitamin d total ii result was > 100 ng/ml. The sample was repeated with a result of > 1000 ng/ml. The vitamin d total ii result using a 1:10 dilution was approximately 8000 ng/ml. This result was reported outside of the laboratory as > 1000 ng/ml where the doctor questioned the result. The customer ran the sample using vitamin d total and the result was 69 ng/ml. The customer repeated the sample using vitamin d total ii and the result was still above the technical range. There was no allegation that an adverse event occurred. The e601 module serial number was (b)(4). The customer stated qc results were within the acceptable range.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01502
MDR Report Key8519019
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-16
Date of Report2019-05-01
Date of Event2019-03-18
Date Mfgr Received2019-03-27
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS VITAMIN D TOTAL II
Generic NameVITAMIN D TEST SYSTEM
Product CodeMRG
Date Received2019-04-16
Model NumberNA
Catalog Number07464215190
Lot Number366696
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.