EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX CDK-1413

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-04-16 for EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX CDK-1413 manufactured by Atricure, Inc..

Event Text Entries

[147621903] (b)(4). The episense device lot number 88524 was returned and evaluated. The complaint was confirmed for zero flow within the flow limiter at the perfusion port.
Patient Sequence No: 1, Text Type: N, H10


[147621904] It was reported that on (b)(6) 2019 a (b)(6) year old male patient underwent a convergent epicardial ablation procedure. Prior to the start of the procedure, the staff tested the device and the saline would not channel through the device when suction was on. When saline was detached from the device, saline would flow without issue. The staff tried flushing the probe with saline through the flush port and it would not flush. A new probe and radio frequency cable were opened, and the procedure was completed with the new device set-up. The procedure was completed with no adverse consequence to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011706110-2019-00019
MDR Report Key8519851
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-04-16
Date of Report2019-04-16
Date of Event2019-03-19
Date Mfgr Received2019-03-19
Device Manufacturer Date2018-11-16
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOHN EHLERT
Manufacturer Street7555 INNOVATION WAY
Manufacturer CityMASON OH 45040
Manufacturer CountryUS
Manufacturer Postal45040
Manufacturer Phone5137554563
Manufacturer G1ATRICURE, INC.
Manufacturer Street7555 INNOVATION WAY
Manufacturer CityMASON OH 45040
Manufacturer CountryUS
Manufacturer Postal Code45040
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX
Generic NameEPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX
Product CodeOCL
Date Received2019-04-16
Returned To Mfg2019-04-08
Model NumberCDK-1413
Catalog NumberCDK-1413
Lot Number88524
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerATRICURE, INC.
Manufacturer Address7555 INNOVATION WAY MASON OH 45040 US 45040


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.