7.5CM ANTERIOR/POSTERIOR RR 750

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-16 for 7.5CM ANTERIOR/POSTERIOR RR 750 manufactured by Arthrocare Corp..

Event Text Entries

[142221369] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[142221370] It was reported that during an unknown procedure the patient was packed bi-laterally with rapid rhino following tumour de-bulking procedure. They don? T know when and if the balloon was inflated. As the patient was waking, they observed that the patient had difficulty breathing. They removed the rapid rhino and observed that it appear to have come apart. The piece in the laryngopharnxy was removed under laryngoscopy. They also removed more material from the nasal airway under endoscopy. Immediate de -gloving of rapid rhino pack with alginate and ending up in laryngeal inlet. Patient could not breath, laryngoscopy was performed to remove alginate from laryngeal inlet, and the nose was packed with bipp (bismuth iodine paraffine paste).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006524618-2019-00190
MDR Report Key8520496
Date Received2019-04-16
Date of Report2019-04-25
Date of Event2019-03-25
Date Mfgr Received2019-04-25
Date Added to Maude2019-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer Phone5123585706
Manufacturer G1ARTHROCARE CORP.
Manufacturer Street7000 W. WILLIAM CANNON
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal Code78735
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name7.5CM ANTERIOR/POSTERIOR
Generic NameBALLOON, EPISTAXIS
Product CodeEMX
Date Received2019-04-16
Catalog NumberRR 750
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARTHROCARE CORP.
Manufacturer Address7000 W. WILLIAM CANNON AUSTIN TX 78735 US 78735


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.