VASCULAR PROBES 7151015

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-19 for VASCULAR PROBES 7151015 manufactured by Synovis Surgical Innovations.

Event Text Entries

[142889160] One single-use device was received for evaluation. A visual inspection was performed and loose particulate matter (pm) was observed on the outside of the inner pouch. Microscopic inspection was performed and the loose pm was determined to be a piece of the suture line. The reported condition was verified. The cause of the condition was determined to be a manufacturing issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[142889161] This report summarizes 1 malfunction events. It was reported that particulate matter (pm) was observed within the inner pouch of a vascular probe. The pm was discovered during incoming inspection and prior to patient use. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2019-02129
MDR Report Key8533779
Date Received2019-04-19
Date of Report2019-04-19
Date Mfgr Received2019-03-31
Date Added to Maude2019-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1SYNOVIS SURGICAL INNOVATIONS
Manufacturer Street2575 UNIVERSITY AVE W
Manufacturer CitySAINT PAUL MN 55114
Manufacturer CountryUS
Manufacturer Postal Code55114
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVASCULAR PROBES
Generic NameDILATOR, VESSEL, SURGICAL
Product CodeDWP
Date Received2019-04-19
Model NumberNA
Catalog Number7151015
Lot NumberSP18H061318448
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerSYNOVIS SURGICAL INNOVATIONS
Manufacturer AddressSAINT PAUL MN


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-19