BD ULTRASAFE? PLUS X100L PNG RFS. 47594130

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-22 for BD ULTRASAFE? PLUS X100L PNG RFS. 47594130 manufactured by Becton Dickinson Hungary Kft (bd).

Event Text Entries

[143044938] Date of event: unknown. A device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[143044939] It was reported there were 200 occurrences where syringes were not able to be assembled correctly as the syringe did not drop into the safety device with a bd ultrasafe? Plus x100l png rfs. The following information was provided by the initial reporter: assembled 500 safety devices. Around 40 % of the syringes can`t be assembled in the standard way. These 40% of syringes did not drop into the safety device. Rather, they get hung up in the device before reaching the snaps. This syringes stick around 1 cm over. The syringe could be easily pushed with a finger completely in the safety device. But this is not the normal assembling process and not common for the clinical production, because we normally have a plunger rod already assembled in this assembling step. If you push the syringe in the safety device by finger you feel a scraping of the syringe in the safety device (area around the spring). During visual inspection afterwards you can`t see any damages or scratches.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009081593-2019-00135
MDR Report Key8536565
Date Received2019-04-22
Date of Report2019-06-04
Date of Event2019-04-01
Date Mfgr Received2019-04-01
Device Manufacturer Date2018-10-15
Date Added to Maude2019-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON DICKINSON HUNGARY KFT (BD)
Manufacturer StreetUVEGGYAR UTCA 3
Manufacturer CityKORNYE TATABANYA 2851
Manufacturer CountryHU
Manufacturer Postal Code2851
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBD ULTRASAFE? PLUS X100L PNG RFS.
Generic NameANTISTICK SYRINGE
Product CodeMEG
Date Received2019-04-22
Catalog Number47594130
Lot Number8288952
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON HUNGARY KFT (BD)
Manufacturer AddressUVEGGYAR UTCA 3 KORNYE TATABANYA 2851 HU 2851


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-22

© 2025 FDA.report