COMPELLA RENT US SCALE PD PPM P7800ARENT01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-22 for COMPELLA RENT US SCALE PD PPM P7800ARENT01 manufactured by Hill-rom Batesville.

Event Text Entries

[142751902] The hill-rom technician found the air supply power cord had been pinched in the side rail, when it was lowered, and needed to be replaced. Per the hill-rom compella bariatric bed system user manual: warning? Before you plug the power cords in, make sure they are not damaged (cuts, exposed wires, worn insulation, etc. ). If either power cord is damaged, do not use the bed. Contact your facility-authorized maintenance person or hill-rom technical support. Warning? Incorrect use or handling of the power cords may cause damage to the power cords. If damage has occurred to either power cord or any of its components, immediately remove the bed from service, and contact your facility-authorized maintenance person or hill-rom technical support. The technician replaced the air supply power cord to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[142751903] Hill-rom received a report from the account stating the power cord for the air supply unit sparked and a flame could be seen. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2019-00173
MDR Report Key8536709
Date Received2019-04-22
Date of Report2019-03-24
Date of Event2019-03-24
Date Mfgr Received2019-03-24
Device Manufacturer Date2014-11-09
Date Added to Maude2019-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOMPELLA RENT US SCALE PD PPM
Generic NameBARIATRIC BED
Product CodeOSI
Date Received2019-04-22
Model NumberP7800ARENT01
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.