PIP SZ. 20 PROXIMAL PIP-200-20P-WW

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-04-22 for PIP SZ. 20 PROXIMAL PIP-200-20P-WW manufactured by Ascension Orthopedics.

Event Text Entries

[142669983] The device was not returned for evaluation, so the failure mode could not be confirmed. A review of the lot records was conducted, and did not discover any indications of problems that could have caused or contributed to the complaint. The reported complaint was not confirmed. While the root cause is unknown for this case, similar issues that resulted in intra-operative pip fractures have been attributed to improperly prepared oblique osteotomy, improper placement of the implant (e. G. , impacting unsupported head with too much force), or improper implant size selection.
Patient Sequence No: 1, Text Type: N, H10


[142669984] An account manager reported that on (b)(6) 2019, the pip-200-20p-ww pip sz. 20 proximal implant head broke off during impaction. The implant stem was stuck inside the bone canal and had to be removed by bending a pair of the hospitals forceps. Minimal debris was left in the surgical site and this was thoroughly checked by the surgeon, and they were able to complete the procedure with a replacement product available. The event lead to 10 minutes surgical delay with no consequences for the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651501-2019-00013
MDR Report Key8537100
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-04-22
Date of Report2019-03-29
Date of Event2019-03-28
Date Mfgr Received2019-03-29
Date Added to Maude2019-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER VIVIAN NELSON
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1ASCENSION ORTHOPEDICS
Manufacturer Street8700 CAMERON ROAD #100
Manufacturer CityAUSTIN TX 78754
Manufacturer CountryUS
Manufacturer Postal Code78754
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePIP SZ. 20 PROXIMAL
Generic NamePIP
Product CodeNEG
Date Received2019-04-22
Catalog NumberPIP-200-20P-WW
Lot Number163401T
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerASCENSION ORTHOPEDICS
Manufacturer Address8700 CAMERON ROAD #100 8700 CAMERON ROAD #100 AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-22

© 2026 FDA.report