TEC 850 1177-9850-SEV

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-22 for TEC 850 1177-9850-SEV manufactured by Datex-ohmeda, Inc..

Event Text Entries

[144848222] For this event, ge healthcare? 's investigation into the reported occurrence is ongoing. A follow-up summary report will be issued when the investigation has been completed. No patient information available. Device evaluation anticipated, but not yet begun.
Patient Sequence No: 1, Text Type: N, H10


[144848223] This report summarizes 1 malfunction event. A review of the event indicated that model 1177-9850-sev vaporizer experienced low output. It was reported that when the unit was set at 4%, it reads at 1. 5% causing the patient to wake up during surgery. There was no report of patient injury or recall. The report was received from a single source. The reported event did involve a patient. There was no patient information available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2112667-2019-00178
MDR Report Key8537431
Date Received2019-04-22
Date of Report2019-06-05
Date Mfgr Received2019-06-03
Device Manufacturer Date1970-01-01
Date Added to Maude2019-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactJOHN SZALINSKI
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTEC 850
Generic NameVAPORIZER
Product CodeCAD
Date Received2019-04-22
Model Number1177-9850-SEV
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDATEX-OHMEDA, INC.
Manufacturer Address3030 OHMEDA DR, MADISON, WI 53718 US 53718


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.