MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-23 for ELECSYS VITAMIN D TOTAL II 07028148190 manufactured by Roche Diagnostics.
[142768912]
This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[142768913]
The initial reporter stated that they received an erroneous high result for one patient sample tested with the elecsys vitamin d total gen. 2 assay on a cobas 8000 e 801 module. The erroneous result was not reported outside of the laboratory. The serum sample initially resulted with a vitamin d value of > 100 ng/ml. The sample was stored frozen and mixed prior to repeating. The sample was diluted times two and repeated, resulting with a value of 24 ng/ml. No adverse events were alleged to have occurred with the patient. The serial number of the e 801 analyzer is (b)(4). Roche diagnostics has issued an urgent medical device correction for this issue, entitled "elecsys? Vitamin d total ii assay? Non-reproducible false high results. " this correction has been reported to fda. During the implementation of the elecsys vitamin d total ii assay on the modular analytics e 170 module, and cobas e 601 and 602 systems, there were reports of non-reproducible, false high results. These customer observations were made in comparisons of duplicate measurements during assay validation of elecsys vitamin d total ii assay or in method comparisons with elecsys vitamin d assay (i. E. , during conversions), where the falsely elevated discrepant value did not fit the expected result. The observed elevated results reported with the elecsys vitamin d ii assay were not confirmed upon repeat testing of the affected samples. The observed findings: the first result is elevated, either above the upper end of the measuring range (>100 ng/ml or >250 nmol/ml), or within the measuring range; repeated results are significantly lower. Rare cases have been reported on the cobas e 411 analyzer and the cobas e 801 module. Roche is conducting investigations into the reported issue and has determined that the elecsys? Vitamin d total ii assay is strongly affected by pre-analytical errors. The investigations have reproduced these findings when requirements for pre-analytical sample handling have not been met for plasma samples. Further investigations into the reported issue are ongoing. As a result, the pre-analytical sample quality and compliance to the tube manufacturer? S specifications is very important to assure a high quality sample in order to minimize the risk of occurrence. Investigations are ongoing to determine the root cause of this issue. A workaround has been provided to customers.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2019-01559 |
MDR Report Key | 8539309 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2019-04-23 |
Date of Report | 2019-06-21 |
Date of Event | 2019-04-03 |
Date Mfgr Received | 2019-04-04 |
Date Added to Maude | 2019-04-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | RES 8008 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ELECSYS VITAMIN D TOTAL II |
Generic Name | VITAMIN D TEST SYSTEM |
Product Code | MRG |
Date Received | 2019-04-23 |
Model Number | NA |
Catalog Number | 07028148190 |
Lot Number | 355684 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-23 |