ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO 06368590190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-23 for ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO 06368590190 manufactured by Roche Diagnostics.

Event Text Entries

[142771217] The event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[142771218] The initial reporter complained of a questionable roche diagnostics cobas elecsys anti-tpo result for 1 patient tested on a cobas e 411 immunoassay analyzer. The erroneous result was reported outside of the laboratory. There was no allegation of an adverse event. The cobas e411 serial number was (b)(4). Following the discrepant results, the customer performed qc testing and the results were not acceptable. After performing a liquid flow cleaning (lfc) maintenance activity, the control results were again acceptable. This behavior is consistent with suboptimal sample conditions. For the cobas e411 used at the investigation site, the serial number was (b)(4). The anti-tpo used on this instrument was lot 368321 with an expiration date of oct-2019. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01557
MDR Report Key8539311
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-23
Date of Report2019-04-29
Date of Event2019-03-14
Date Mfgr Received2019-04-04
Date Added to Maude2019-04-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
Generic NameSYSTEM, TEST, THYROID AUTOANTIBODY
Product CodeJZO
Date Received2019-04-23
Model NumberNA
Catalog Number06368590190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-23

© 2026 FDA.report