BODY WEIGHT SUPPORT LIGHT RB011

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-23 for BODY WEIGHT SUPPORT LIGHT RB011 manufactured by Motek Medical B.v..

Event Text Entries

[143033800] Detailed and precise information about event and patient information from customer is still missing. We have attempted to get additional information from customer several times via email and phone, but until now, information was not provided. Further analysis according to inspection details are therefore ongoing. Root cause analysis is not completed.
Patient Sequence No: 1, Text Type: N, H10


[143033801] During a treatment session (gait rehabilitation), the yoke (attachment point of the patient harness) of the medical device (body weight support (bws) light) came loose and fell off. The patient did not get injured. Possible risk is that the yoke during treatment can detach and fall to the patient head from distance of about 15 cm, which can be very dangerous for specific patient population (patients after recent craniotomy/craniectomy surgery). It is important to note that patient harness is also attached to a safety line to prevent patients from falling, and this safety system was not affected and stayed operational throughout.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006750942-2019-00001
MDR Report Key8540056
Date Received2019-04-23
Date of Report2019-04-23
Date of Event2019-03-27
Date Mfgr Received2019-03-27
Date Added to Maude2019-04-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKEVIN WALLS
Manufacturer Street33 GOLDEN EAGLE LANE
Manufacturer CityLITTLETON CO 80127
Manufacturer CountryUS
Manufacturer Postal80127
Manufacturer Phone7209625412
Manufacturer G1MOTEK MEDICAL B.V.
Manufacturer StreetHOGEHILWEG 18 UNIT C
Manufacturer CityAMSTERDAM, NOORD-HOLLAND 1101CD
Manufacturer CountryNL
Manufacturer Postal Code1101CD
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBODY WEIGHT SUPPORT LIGHT
Generic NamePOWERED EXERCISER
Product CodeBXB
Date Received2019-04-23
Model NumberRB011
Catalog NumberRB011
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMOTEK MEDICAL B.V.
Manufacturer AddressHOGEHILWEG 18 UNIT C AMSTERDAM, NOORD-HOLLAND 1101CD NL 1101CD


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-04-23