POWERED CHAIR/TRANSPORT CHAIR 54PFS04-TS7 54P SERIES RECLINER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-23 for POWERED CHAIR/TRANSPORT CHAIR 54PFS04-TS7 54P SERIES RECLINER manufactured by Champion Manufacturing.

Event Text Entries

[142811722] Repair history at this facility was reviewed. No repairs to any chairs for anything related to foot trays were found.
Patient Sequence No: 0, Text Type: N, H10


[142811723] On 3/26/19 phone call from facility to get written permission to remove foot trays from their chairs, due to several near falls (over the time they have had their chairs) where nurses had to catch patients. The patients think they are standing on the ground, but are standing on the foot tray, when they take a step the stumble or fall. At the time of this phone call, the facility was made aware of the optional yellow foot tray with caution label stating: stow when not in use. Use with staff assistance only. Nurse at the facility provided details of the incident. Patient finished treatment, stood up on foot tray and was waving to someone out the window. She took a step and fell to the floor, fracturing her trochanter head (which the nurse noted was the top of the femur that connects into the hip joint). This required surgery. The nurse also noted that since then the patient has passed, due to the disease process, not the fall.
Patient Sequence No: 0, Text Type: D, B5


MAUDE Entry Details

Report Number1834066-2019-00001
MDR Report Key8540630
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-23
Date of Report2019-04-22
Date of Event2019-02-11
Date Mfgr Received2019-03-26
Device Manufacturer Date2015-02-26
Date Added to Maude2019-04-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINE MILLER
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal46516
Manufacturer Phone8009985018
Manufacturer G1CHAMPION MANUFACTURING INC
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal Code46516
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOWERED CHAIR/TRANSPORT CHAIR
Generic NameCHAIR, SURGICAL AC-POWERED
Product CodeGBB
Date Received2019-04-23
Model Number54PFS04-TS7
Catalog Number54P SERIES RECLINER
Lot NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHAMPION MANUFACTURING
Manufacturer Address2601 INDUSTRIAL PARKWAY ELKHART IN 46516 US 46516


Patients

Patient NumberTreatmentOutcomeDate
001. Required No Informationntervention 2019-04-23