EXABALTE 4000 SYS940350

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-04-24 for EXABALTE 4000 SYS940350 manufactured by Insightec Ltd.

Event Text Entries

[142878085] The system technical performance during treatment was thoroughly reviewed;no technical issues were found. No system malfunction occurred. All sonication parameters were within the normal ranges with no exceptional usage.
Patient Sequence No: 1, Text Type: N, H10


[142878086] The patient underwent exablate thalamotomy for essential tremor and no side effects were recorded on the day of treatment. On (b)(6) 2018, insightec learned that after treatment the patient was a little unsteady on her feet, but she managed to walk from the mr room. After returning to the ward patient had difficulties in walking. The patient was hospitalized for 2 weeks, being treated with a reducing dose of dexamethasone for 3 days. The patient through this time remained unstable but was able to walk with aids (frame). On (b)(6) 2018, when the case was discussed with the site team, the patient continues to slowly improve and is receiving physiotherapy. At this time, there is no additional information.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615058-2018-00006
MDR Report Key8542913
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-04-24
Date of Report2018-10-08
Date of Event2018-08-24
Date Mfgr Received2018-09-07
Device Manufacturer Date2017-12-11
Date Added to Maude2019-04-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR IDAN COHEN
Manufacturer Street5 NACHUM HETH STREET
Manufacturer CityTIRAT CARMEL, 31290
Manufacturer CountryIS
Manufacturer Postal31290
Manufacturer G1INSIGHTEC LTD
Manufacturer Street5 NACHUM HETH STREET
Manufacturer CityTIRAT CARMEL, 31290
Manufacturer CountryIS
Manufacturer Postal Code31290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXABALTE 4000
Generic NameMR GUIDED FOCUSED ULTRASOUND SYSTEM
Product CodePOH
Date Received2019-04-24
Model Number4000
Catalog NumberSYS940350
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINSIGHTEC LTD
Manufacturer Address5 NACHUM HETH STREET TIRAT CARMEL, 31290 IS 31290


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Deathisabilit 2019-04-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.