TOSOH HLC-723G8 ANALYZER G8 021560

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-24 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Bioscience, Inc..

Event Text Entries

[147062388] The field service engineer (fse) evaluated the g8 analyzer. The fse replaced several valves and left the analyzer operational. The valves were returned to the tosoh instrument service center (isc). Isc could not duplicate the failures in the reported events.
Patient Sequence No: 1, Text Type: N, H10


[147062389] This report summarizes 1 malfunction event for the g8 analyzer. The review of the event indicated that the g8 analyzer experienced pressure low error messages, which caused delayed reporting of critical patient results. There was no indication of patient intervention or adverse health consequences due to the delayed reporting in patient results. This event did not necessitate remedial action to prevent an unreasonable risk of substantial harm to the public health.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005529799-2019-00080
MDR Report Key8543683
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-24
Date of Report2019-04-24
Date Added to Maude2019-04-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS. DORIA ESQUIVEL
Manufacturer Street6000 SHORELINE COURT STE 101
Manufacturer CitySOUTH SAN FRANCISCO CA 94080
Manufacturer CountryUS
Manufacturer Postal94080
Manufacturer Phone6506368123
Manufacturer G1TOSOH BIOSCIENCE, INC.
Manufacturer Street3600 GANTZ ROAD
Manufacturer CityGROVE CITY OH 43123
Manufacturer CountryUS
Manufacturer Postal Code43123
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSOH HLC-723G8 ANALYZER G8
Generic NameG8
Product CodeLCP
Date Received2019-04-24
Model NumberG8
Catalog Number021560
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerTOSOH BIOSCIENCE, INC.
Manufacturer Address3600 GANTZ ROAD GROVE CITY OH 43123 US 43123


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-24

© 2026 FDA.report