Event Text Entries

[142931584] The customer evaluated the analyzer with the assistance of field service engineer (fse) and determined that the customer had been installing the filter incorrectly. The customer reinstalled the filter correctly and the pressure was in normal range. The analyzer was operating as expected. There was no further action required by fse.
Patient Sequence No: 1, Text Type: N, H10


[142931585] This report summarizes 1 malfunction event on the g8 analyzer. The review of the event indicated that the g8 analyzer experienced pressure error messages, which caused delayed reporting of critical patient results. This report from was received from one source. There was no indication of patient intervention or adverse health consequences due to the delayed reporting in patient results. This event did not necessitate remedial action to prevent an unreasonable risk of substantial harm to the public health.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8031673-2019-00085
MDR Report Key8543686
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-24
Date of Report2019-04-24
Date Added to Maude2019-04-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS. DORIA ESQUIVEL
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO 1058623
Manufacturer CountryJA
Manufacturer Postal1058623
Manufacturer G1TOSOH CORPORATION
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO 1058623
Manufacturer CountryJA
Manufacturer Postal Code1058623
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSOH HLC-723G8 ANALYZER G8
Generic NameG8
Product CodeLCP
Date Received2019-04-24
Model NumberG8
Catalog Number021560
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerTOSOH CORPORATION
Manufacturer AddressSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO 1058623 JA 1058623


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-24