MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-24 for MICRODRILL SERIES STRAIGHT ATTACHMENT 5100015250 manufactured by Stryker Instruments-kalamazoo.
[143082449]
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program. Reported events: 15? Events were reported for this quarter. Product return status: 11 devices were received. 1 device was not available for evaluation. 3 device investigation types have not yet been determined. ? Event confirmation status: 5 reported events were confirmed; the cause traced to component failure. 5 reported events were not confirmed. 1 device evaluation is still in progress. Evaluation results: 4 devices were found to be affected by internal component corrosion. 2 devices were found to be affected by worn internal components. 1 device was found to be affected by lubrication breakdown. 3 devices were found to be affected by components covered in debris. 1 device was found to be affected by shattered bearings. Additional information: 15? Devices were not labeled for single-use. 15 devices were not reprocessed and reused.
Patient Sequence No: 1, Text Type: N, H10
[143082450]
This report summarizes 15 malfunction events in which the device reportedly overheated. 12 events had no patient involvement; no patient impact. 1 event had patient involvement; no patient impact. 2 events had the patient receive a burn.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0001811755-2019-01410 |
MDR Report Key | 8546243 |
Date Received | 2019-04-24 |
Date of Report | 2019-07-26 |
Date of Event | 2019-01-01 |
Date Mfgr Received | 2019-01-01 |
Date Added to Maude | 2019-04-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. ZACH BAKER |
Manufacturer Street | 4100 EAST MILHAM AVENUE |
Manufacturer City | KALAMAZOO MI 49001 |
Manufacturer Country | US |
Manufacturer Postal | 49001 |
Manufacturer Phone | 2693237700 |
Manufacturer G1 | STRYKER INSTRUMENTS-KALAMAZOO |
Manufacturer Street | 4100 EAST MILHAM AVENUE |
Manufacturer City | KALAMAZOO MI 49001 |
Manufacturer Country | US |
Manufacturer Postal Code | 49001 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | MICRODRILL SERIES STRAIGHT ATTACHMENT |
Generic Name | UNIT, OPERATIVE DENTAL |
Product Code | EIA |
Date Received | 2019-04-24 |
Catalog Number | 5100015250 |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER INSTRUMENTS-KALAMAZOO |
Manufacturer Address | 4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-24 |